MSI MS-9652 Manuel d'utilisateur Page 122

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 134
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 121
Report No 088213R-ITCEP07V04
Page: 119 of 131
Product Notebook
Test Item Power frequency magnetic field
Test Mode Mode 2
Date of Test 2008/08/26 Test Site No.3 Shielded Room
Polarization
Frequency
(Hz)
Magnetic
Strength
(A/m)
Required
Performance
Criteria
Performance
Criteria
Complied To
Test Result
X Orientation 50 1 A A PASS
Y Orientation 50 1 A A PASS
Z Orientation 50 1 A A PASS
Meet criteria A: Operate as intended during and after the test
Meet criteria B: Operate as intended after the test
Meet criteria C: Loss/Error of function
Additional Information
EUT stopped operation and could / could not be reset by operator at kV
of Line
.
No false alarms or other malfunctions were observed during or after the test. The
acceptance criteria were met, and the EUT passed the test.
Vue de la page 121
1 2 ... 117 118 119 120 121 122 123 124 125 126 127 ... 133 134

Commentaires sur ces manuels

Pas de commentaire